Test No. 115: Surface Tension of Aqueous Solutions

This Test Guideline describes methods to determine the surface tension (in N/m) of aqueous solutions. The methods are based on the measurement of the force which it is necessary to exert vertically on a stirrup or ring, in contact with the surface of the liquid, in order to separate it from the surface, or on a plate, with an edge in contact with the surface, in order to draw up the film that has formed.

There are four different methods: the plate method, the stirrup method, the ring method and the OECD harmonized ring method. They are described in detail in the ISO Standard 304-1985. The methods described are applicable to aqueous solutions of most substances regardless of their degree of purity. The concentration should be 90% of the saturation solubility, but must below 1g/l. This shall therefore be carried out under a protective cover to avoid interference at 20°C approximately. The ring is immersed below the surface of the solution. Then the table top, where the measurement vessel is placed, is lowered gradually and evenly at a rate of approximately 0.5 cm/min to detach the ring from the surface until the maximum force is reached. The force is read on the tensiometer. After completing the first measurement, measurements are repeated until a constant surface tension value is reached.

27 Jul 1995 7 pages English Also available in: French

https://doi.org/10.1787/9789264069787-en 9789264069787 (PDF)

Author(s): OECD